3 Remote Interface Reference
Output Settings and Operation Commands
86 E364xA User’s and Service Guide
INSTrument[:SELect]
{OUTPut1|OUTPut2|OUT1|OUT2}
This command selects the output to be programmed by the
output identifier. The outputs of the power supply are
considered as two logical instruments. The INSTrument
command provides a mechanism to identify and select an
output.
When one output is selected, the other output is unavailable
for programming until selected. The commands which are
affected by the INSTrument command are the output setting
command (SOURce), measurement command (MEASure), and
calibration command (CALibration).
INSTrument[:SELect]?
Return the currently selected output by the
INSTrument[:SELect] or INSTrument:NSELect
command. The returned value is OUTP1 or OUTP2.
INSTrument:NSELect {1|2}
This command selects the output to be programmed by a
numeric value instead of the output identifier used in the
INSTrument:NSELect or INSTrument[:SELect]
command. 1 selects output1, and 2 selects output2.
INSTrument:NSELect?
Return the currently selected output by the
INSTrument[:SELect]? or INSTrument[:NSELect]?
command. The returned parameter is 1 for output1, and 2
for output2.
CURR:STEP 0.02 Set the step size to 0.02 A.
CURR DOWN Decrease the output current.