Tutorial 6
Remote Programming
E364xA User’s and Service Guide 167
Since up- programming speed is aided by the conduction of
the series regulating transistor, while down- programming
normally has no active element aiding in the discharge of
the output capacitor, laboratory power supplies normally
program upward more rapidly than downward.