Triggers 10
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide 159
Setup and Hold Trigger
The Setup and Hold trigger looks for setup and hold violations.
One oscilloscope channel probes the clock signal and another channel probes the
data signal.
To trigger on setup & hold violations:
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to
select Setup and Hold.
3 Press the Clock softkey; then, turn the Entry knob to select the input channel
with the clock signal.
4 Set the appropriate trigger tevel for the clock signal using the Trigger Level
knob.
5 Press the Rising Edge or Falling Edge softkey to specify the clock edge being used.
6 Press the Data softkey; then, turn the Entry knob to select the input channel
with the data signal.
7 Set the appropriate trigger tevel for the data signal using the Trigger Level
knob.
8 Press the <Setup softkey, and turn the Entry knob to select the setup time.
Data
Clock
(rising
edge)
Setup time Hold time

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision MSO-X 3014A and is the answer not in the manual?

Keysight Technologies InfiniiVision MSO-X 3014A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision MSO-X 3014A
CategoryTest Equipment
LanguageEnglish

Related product manuals