282 Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
18 Save/Recall (Setups, Screens, Data)
To recall setup files
Setup files can be recalled from one of 10 internal (\User Files) locations or from
an external USB storage device.
1 Press [Save/Recall] > Recall > Recall:; then, turn the Entry knob to select Setup
(*.scp).
2 Press the softkey in the second position and use the Entry knob to navigate to
the file to recall. See "To navigate storage locations" on page 280.
3 Press the Press to Recall softkey.
A message indicating whether the recall was successful is displayed.
4 If you would like to clear the display, press Clear Display.
To recall mask files
Mask files can be recalled from one of four internal (\Keysight Flash) locations or
from an external USB storage device.
1 Press [Save/Recall] > Recall > Recall:; then, turn the Entry knob to select Mask
(*.msk).
2 Press the softkey in the second position and use the Entry knob to navigate to
the file to recall. See "To navigate storage locations" on page 280.
3 Press the Press to Recall softkey.
A message indicating whether the recall was successful is displayed.
4 If you would like to clear the display or clear the recalled mask, press Clear
Display or Clear Mask.
To recall reference waveform files from a USB storage device
1 Press the [Save/Recall] key.
2 In the Save/Recall Menu, press the Recall softkey.
3 In the Recall Menu, press the Recall softkey and turn the Entry knob to select
Reference Waveform data (*.h5) .
4 Press the To Ref: softkey and turn the Entry knob to select the desired reference
waveform location.
5 Press the softkey in the second position and use the Entry knob to navigate to
the file to recall. See "To navigate storage locations" on page 280.
6 Press the Press to Recall softkey.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision MSO-X 3014A and is the answer not in the manual?

Keysight Technologies InfiniiVision MSO-X 3014A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision MSO-X 3014A
CategoryTest Equipment
LanguageEnglish

Related product manuals