EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 3014A

Keysight Technologies InfiniiVision MSO-X 3014A
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
239
Keysight InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
15 Mask Testing
To create a mask from a "golden" waveform (Automask) / 239
Mask Test Setup Options / 241
Mask Statistics / 244
To manually modify a mask file / 245
Building a Mask File / 248
One way to verify a waveform's compliance to a particular set of parameters is to
use mask testing. A mask defines a region of the oscilloscope's display in which
the waveform must remain in order to comply with chosen parameters.
Compliance to the mask is verified point-by-point across the display. Mask test
operates on displayed analog channels; it does not operate on channels that are
not displayed.
Mask test is a license-enabled feature.
To create a mask from a "golden" waveform (Automask)
A golden waveform meets all chosen parameters, and it is the waveform to which
all others will be compared.
1 Configure the oscilloscope to display the golden waveform.
2 Press the [Analyze] key.
3 Press Features; then, select Mask Test.
4 Press Features again to enable mask testing.

Table of Contents

Related product manuals