222 Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
14 Measurements
Preshoot
Preshoot is distortion that precedes a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being measured (edge
closest to the trigger reference point).
Average
Average is the sum of the levels of the waveform samples divided by the number
of samples.
Where x
i
= value at ith point being measured, n = number of points in
measurement interval.
The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an integral
number of periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show what interval of the waveform is being measured.
Rising edge preshοot =
local Maximum − D Top
Amplitude
× 100
Falling edge preshοot =
Base − D local Minimum
Amplitude
× 100
Preshoot
local Maximum
local Minimum
Base
Top
Preshoot
Average =
∑
x
i
n

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision MSO-X 3014A and is the answer not in the manual?

Keysight Technologies InfiniiVision MSO-X 3014A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision MSO-X 3014A
CategoryTest Equipment
LanguageEnglish

Related product manuals