Measurement
97
All Channel
Sets active channel trigger mode to hold.
Active Channel
Triggers Active channel alone.
Sweep with unlimited times (continuous sweep)
1. Press Trigger > Continuous to select continuous sweep.
When continuous sweep is selected, sweeping can be
repeated unlimited times after it is triggered. Press the Hold
key to stop the sweeping. No further sweeping can be triggered
after it is stopped.
Stopping sweep
1. Press Trigger > Hold to stop the sweep.
Pressing the Hold key immediately stops sweeping,
regardless of which sweeping state (single, specified times, or
continuous) is selected. No further sweeping can be triggered
after it is stopped.
Sweeping Multiple Sweep Ranges with Different Conditions in a Single Action
(Segment Sweep)
Concept of Segment Sweep
To perform a segment sweep, you must define two or more frequency
ranges, called segments, and then specify the number of points, OSC level,
Bias Level, measurement time, sweep delay time, and sweep time for each
segment. All segments are swept sequentially as if swept in one sweep
operation.
By skipping the frequency range, which does not need to
be measured, you can sweep and measure only the portions you
need.
You can define the optimum measurement conditions for
each of the segments you designate. For example, you can specify as
many points as possible in a segment requiring high trace resolution
and as few points as possible in a segment not requiring high
resolution. This shortens the measurement time, enabling you to
optimize the overall measurement throughput by not having to
perform the entire operation under the same measurement conditions
of a particular frequency range.
Preparing Segment Sweep Table
Set the segment sweep with the segment sweep table.