Programming
321
0 - 3 Not used Always 0
4 Measurement Set to "1" during measurement
5
Waiting for
Trigger
Set to "1" while the instrument is
waiting for a trigger.
6 - 15 Not used Always 0
Issuing the *CLS command clears all bits from the operation status event
register.
Status Bit Definitions of the Questionable Status Condition Register
Bit
Position
Name Description
0 - 9 Not used Always 0
10
Limit Test Fail
(Questionable limit
status register
summary)
Set to "1" while one of the
enabled bits in the
questionable limit status event
register is set to "1."
11 - 15 Not used Always 0
Status Bit Definitions of the Questionable Status Event Register
Bit
Position
Name Description
0 - 9 Not used Always 0
10
Limit Test Fail
(Questionable limit
status register
summary)
Set to "1" when a transition of
the condition register occurs if
the transition filters are set as
valid values.
11 - 15 Not used Always 0
Issuing the *CLS command clears all bits from the questionable status
event register.