Programming
379
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.LIMIT.REPORT.ALL
:CALC<Ch>:LIM:REP:MARK<Mk>
Type
Query
Syntax
:CALCulate<Ch>[:SELected]:LIMit:REPort:MARKer<Mk>[:ALL]?
Description
This command returns the limit test result of measurement point of the
active marker position. (Mk: marker 1 to 9 and reference marker 10.)
If the marker function is off, executing this command causes an error and
invalid values are read out.
If marker discrete is set at Off and the marker is located between two
points, the value at the closest point is returned.
Variable
Query Response
{numeric 1},{numeric 2},{numeric 3},{numeric 4}<newline><^END>
{numeric 1}: Sweep parameter value at Marker position
{numeric 2}: Limit test result
0: FAIL
1: PASS
-1: Limit test was off
{numeric 3}: Upper border value of the limit (0 is returned if no limit line
is set.)
{numeric 4}: Lower border value of the limit (0 is returned if no limit line
is set.)
Examples
Dim Var as Variant
Ana.WriteString ":CALC1:LIM:REP:MARK1?", True
Var=Ana.ReadList(ASCIIType_R8, ",")
Related Commands
:CALC:LIM