E4990A Help
380
:CALC:LIM:REP:ALL
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.LIMIT.REPORT.MARKER(Mk).ALL
:CALC<Ch>:LIM:REP:POIN
Type
Query
Syntax
:CALCulate<Ch>[:SELected]:LIMit:REPort:POINts?
Description
This command returns the number of the failed points of the limit test.
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Long
Ana.WriteString ":CALC1:LIM:REP:POIN?", True
Var=Ana.ReadNumber
Related Commands
:CALC:LIM
:CALC:LIM:REP
Equivalent Softkey
No equivalent key is available on the front panel.
COM Command Syntax (Internal Control Only)
SCPI.CALCULATE(Ch).SELECTED.LIMIT.REPORT.POINTS
:CALC<Ch>:MARK<Mk>
Type
Command/Query
Syntax
:CALCulate<Ch>[:SELected]:MARKer<Mk>[:STATe] {ON|OFF|1|0}
:CALCulate<Ch>[:SELected]:MARKer<Mk>[:STATe]?
Description
This command turns ON/OFF the display of marker 1 to 9 (Mk) and
reference marker (Mk:10).
Variable
Parameter Selection Option