Programming
407
:CALCulate<Ch>[:SELected]:MARKer<Mk>:SET
{CENTer|STARt|STOP|RLEVel|ZOOM}
Description
This command sets the value at the position of marker 1 to 9 (Mk) and
reference marker (Mk:10) to the value of the stimulus.
Variable
Parameter Selection Option
Description Type of stimulus value
Data Type Character string type (String)
Range
"STARt": Sets the sweep start value to the stimulus
value at the marker position.
"STOP": Sets the sweep stop value to the stimulus value
at the marker position.
"CENTer": Sets the sweep center value to the stimulus
value at the marker position.
"RLEVel": Sets the reference line value to the response
value at the marker position.
"ZOOM": Sets the span value in order to zoom the span
by specified zoom aperture.
Preset
Value
-
Examples
Dim Var as String
Var= "CENT"
Ana.WriteString ":CALC1:MARK1:SET " & Var, True
Related Commands
:CALC:PAR:SEL
:CALC:MARK:REF
:CALC:MARK:ZAP
Equivalent Softkey
Marker Fctn > Marker > Start
Marker Fctn > Marker > Stop
Marker Fctn > Marker > Center
Marker Fctn > Marker > Reference