E4990A Help
476
This command can be used to set/get the format data using the following
SCPI commands:
:CALC:DATA:FDAT
:CALC:DATA:FMEM
:CALC:DATA:RDAT
:CALC:DATA:RMEM
:CALC:DATA:XAX
:CALC:FUNC:DATA
:CALC:LIM:DATA
:CALC:LIM:REP
:CALC:LIM:REP:ALL
:SENS:FREQ:DATA
:SENS:SEGM:DATA
Variable
Parameter Selection Option
Description Data transfer mode
Data Type Character string type (String)
Range
ASCii: ASCII transfer format
REAL: IEEE 64-bit floating point binary transfer format
REAL32: IEEE 32-bit floating point binary transfer
format
Preset
Value
ASCii
Query Response
{ASC|REAL|REAL32}<newline><^END>
Examples
Dim Var as String
Var= "ASC"
Ana.WriteString ":FORM:DATA " & Var, True
Ana.WriteString ":FORM:DATA?", True
Var=Ana.ReadString