Description Point averaging On/Off state
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value
OFF
Query Response
{1|0}<newline><^END>
Examples
Dim Var as Integer
Ana.WriteString ":SENS1:AVER ON", True
Ana.WriteString ":SENS1:AVER?", True
Var=Ana.ReadNumber
Related Commands
:SENS:AVER:COUN
Equivalent Softkey
Average > Point Avg
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).AVERAGE.STATE
:SENS<Ch>:AVER:COUN
Type
Command/Query
Syntax
:SENSe<Ch>:AVERage:COUNt <Value>
:SENSe<Ch>:AVERage:COUNt?
Description
This command sets/gets the point averaging factor.
Variable
Parameter Value
Description Averaging factor
Data Type Numeric type (Integer)
Range 1 ~ 999