Programming
519
Data Type Numeric type (Real)
Range -1M ~ 1M
Preset Value
0
Unit H
Resolution 1p
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= -1000000
Ana.WriteString ":SENS1:CORR1:CKIT:LOAD:L " & Str(Var), True
Ana.WriteString ":SENS1:CORR1:CKIT:LOAD:L?", True
Var=Ana.ReadNumber
Related Commands
:SENS:CORR:CKIT:LOAD:MOD
Equivalent Softkey
Cal > User Cal > Define Value > Load Induct (L)
Cal > Fixture Compen > Define Value > Load Induct (L)
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).CORRECTION(Type).CKIT.LOAD.L
:SENS<Ch>:CORR<Type>:CKIT:LOAD:MOD
Type
Command/Query
Syntax
:SENSe<Ch>:CORRection<Type>:CKIT:LOAD:MODel {EQUiv|TABLe}
:SENSe<Ch>:CORRection<Type>:CKIT:LOAD:MODel?
Description
This command selects the load standard impedance model for the user
calibration and fixture compensation.
(<type>: User calibration=1, Fixture Compensation=2)
Variable
Parameter Selection Option