E4990A Help
520
Description Load standard impedance model
Data Type Character string type (String)
Range
EQUiv: Equivalent Circuit
TABLe: f-Z Table
Preset Value
EQUiv
Query Response
{EQU|TABL}<newline><^END>
Examples
Dim Var as String
Var= "EQU"
Ana.WriteString ":SENS1:CORR1:CKIT:LOAD:MOD " & Var, True
Ana.WriteString ":SENS1:CORR1:CKIT:LOAD:MOD?", True
Var=Ana.ReadString
Related Commands
:SENS:CORR:CKIT:LOAD:TABL
:SENS:CORR:CKIT:LOAD:C
:SENS:CORR:CKIT:LOAD:L
:SENS:CORR:CKIT:LOAD:R
Equivalent Softkey
Cal > User Cal > Define Value > Load Model > Equivalent Circuit | f-Z Table
Cal > Fixture Compen > Define Value > Load Model > Equivalent Circuit | f-Z
Table
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).CORRECTION(Type).CKIT.LOAD.MODEL
:SENS<Ch>:CORR<Type>:CKIT:LOAD:R
Type
Command/Query
Syntax
:SENSe<Ch>:CORRection<Type>:CKIT:LOAD:R <Value>
:SENSe<Ch>:CORRection<Type>:CKIT:LOAD:R?
Description
This command sets/gets the load resistance value for the user calibration
and fixture compensation.