6600 Product Manual
Operator Manual
11
EN 61000-3-3
use by healthcare professionals only. The models
6600 series may cause radio interference or may
disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as
re-orienting or relocating the 6600 series or
shielding the location.
Electromagnetic immunity
Guidance and manufacturer’s declaration electromagnetic immunity
The models 6600 series are intended for use in the electromagnetic environment
specified below. The customer or the user of the 6600 should assure that it is used in
such an environment.
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic
environment – guide
Electrostatic
discharge (ESD)
EN 61000-4-2
±8kV contact
±2kV, ±4kV, ±8kV,
±15kV air
±8kV contact
±2kV, ±4kV, ±8kV,
±15kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast
transient/burst EN
61000-4-4
±2kV for power
supply lines
±1kV for signal
input/output lines
±2kV for power
supply lines
±1kV for signal
output lines
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Surge
EN 61000-4-5
±1kV line to line
±2kV line to earth
±1kV line to line
±2kV line to earth
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
<5% UT
(>95% dip in UT) For
0.5 cycle
<5% UT
(>95% dip in UT)
for 1 cycles
<5% UT
(>95% dip in UT)
For 0.5 cycle
<5% UT
(>95% dip in UT)
for 1 cycles
Mains power
quality should be
that of a typical
commercial or
hospital
environment. If the