6600 Product manual
Operator Manual
12
EN 61000-4-11
70% UT
(30% dip in UT)
for 25/30 cycles
<5% UT
(>95% dip in UT)
for 5/6 sec
70% UT
(30% dip in UT)
for 25/30 cycles
<5% UT
(>95% dip in UT)
for 5/6 sec
user of the models
6600 series requires
continued operation
during power mains
interruptions, it is
recommended that
the models 6600
series be powered
from an
uninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
Magnetic field EN
61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
U
T:
AC network voltage before applying test voltage
Guide and manufacturer's statement – electromagnetic immunity
(Equipment or system) is intended to be used in the electromagnetic environment
specified below, and the purchaser or user of (device or system) should ensure that it is
used in this electromagnetic environment:
Immunity test
EN 60601 Test level
Complain
CE level
Electromagnetic environment –
guide
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3Vrms
0.15MHz to 80MHz
6 Vrms in ISM
bands
3V/m
3Vrms
0.15MHz
to 80MHz
6 Vrms in
ISM bands
Portable and mobile RF
communications equipment
should be used no closer to any
part of the models 6600 series,
including cables, than the
recommended separation

Table of Contents