remove the connecting plate between the tube holding hoops and the control panel.
2. Take the tube out of the packing box and place it to the hoop horizontally, and adjust
its position so that it is just stuck inside the hoop.
CAUTION: X-ray tube is vulnerable parts, and great attention shall be paid in the
course of transport and installation, so as to avoid the fall of the tube from
damaging.
NOTE: When installing the tube, the tube outlet shall be downward, and the
cable connection port shall be towards the column direction.
3. Assemble the other half of the hoop and control panel connection plate taken out in
Step 1, and lock the hoop with the four hexagon head screws so as to fix the tube.
4. After the operations above are completed, install the high-voltage cable joint of the
tube.
Please refer to the section Installation Instructions of “X-ray Tube" for more details.
4.1.3.5 Installation of Collimator and Wiring
Please refer to the section Installation Instructions of “Collimator" for more details.
4.1.3.6 Installation of Control Panel Assembly
1. Take the control panel assembly out of the packing box, and align the six screw
holes of the side of base connection plate with the holes of the two sides of the unit
connection plate.
2. Fix the unit assembly with the supplied six stationary screws.
3. Insert the control line inside the corrugated pipe under the tube arm into the socket
on the rear of the control panel.
CAUTION: During installing the unit assembly, be careful not to scratch the
touch screen of the unit control panel.

Table of Contents