Page 9
©2013 Technical Marine Service, Inc. LC-100 V2.97
Features and Definitions
The following are definitions and explanations of LevelCom 100 features and the terminology
used in this manual and the LevelCom 100 configuration interface.
Alarm Setpoint: A user defined reference point that establishes the depth at which a
particular alarm will activate.
Bubbler: A bubbler is a remote sensing liquid level gauge that operates by measuring the
pressure required to blow bubbles from the bottom of a pipe mounted in a tank. If the
specific gravity of the liquid is known, this pressure can be converted to depth of the end of
the pipe below the liquid surface.
Bubbler Pipe: This refers to the pipe installed in the tank and connected to the LevelCom
100. To sense depth the LevelCom 100 blows bubbles from the bottom of this pipe. The
Bubbler to Bottom parameter is the distance between the lower end of the bubbler pipe and
the bottom of the tank.
Buttons: The LevelCom 100 has a set of buttons built into the face plate. These are used
to enter the configuration information into the device, and to locally acknowledge alarm
conditions. In normal operation the buttons may be used to display alternate units, for
example changing from a depth default to read volume or weight. Also the current units may
be read.
Calibrate: The LevelCom 100 is shipped fully calibrated and while in operation automatically
performs a periodic “hardware zero”. Should the unit require field calibration please
reference the calibration procedure beginning on page 72.
Communication: The LevelCom 100 can be purchased with optional communication
interfaces that allow a group of LevelCom 100s to be networked to a central control or data
acquisition computer. These interfaces include RS-422/RS-485 using standard Modbus RTU
Protocol. The machine may also be equipped with interfaces for DeviceNet, or Profibus
communication networks.
Configuration: The LevelCom 100 program allows users to configure the unit to a wide
variety of situations. The unit is shipped with a default configuration and the user adjusts the
parameters to match his own requirements thereby establishing the machine configuration.
Demand: The user may configure the LevelCom 100 to perform sample and/or purge
functions on demand in lieu of automatic timed cycles. This feature is normally used in areas
where the quantity of compressed air is limited or in applications not requiring constant
monitoring.
Hardware Zero: The LevelCom 100 periodically checks the sensor for drift of the zero point
by measuring atmospheric pressure. This value is used to correct the raw pressure readings