Soak Temperature Enter the temperature at which the sample is to
soak while gas flows through the heating station.
Range: 20 to 450
o
C
Default: 30
o
C
Temperature
Ramp Rate
Enter the rate at which the temperature will
change when advancing to the soak temperature.
Range: 5 to 20
o
C per minute
Default: 10
o
C per minute
Soak Time Enter the amount of time the sample will soak.
Range: 0 to 10,000 minutes
Default: 10 minutes
Select a row (highlight the row), then press this
button to insert a row above the one selected.
Select a row (highlight the row), then press this
button to delete the row from the conditions table.
Press this button to clear the conditions table.
The rows do not have to be selected.
Press this button to save the information you
entered.
Press this button to close the Degas Conditions
dialog box.
Insert
Delete
Clear
Save
Close