150
CYCLE DETAILS:
-Consumption: 5ml/cycle
-Cycle time: 80 seconds
To manage liquid expiration, see chapter 2.6.5.
The following figure shows the space of the two extra bottles.
MAINTENANCE
If necessary, you may clean the external body with a moist cloth and non-corrosive substances.
Make sure that the air vents located on the sides of the device are always clear and clean them regularly by
removing any foreign material.
Before replacing with new decontaminant, it is recommended to run three machine cycles adding
demineralized water to the rinsed and clean tanks after having removed any possible residue of hydrogen
peroxide from them.
When changing the liquids, delicately probe the hole between the two golden pins at the end of the liquid
exhaust tube of the tank with an insulin needle (25 G), injecting at least 1 ml of distilled water inside. Do not
use needles thicker than 0.5 mm, to avoid damaging or enlarging the external hole.
5.32. SUPRASPOR
CODE 109350
Peroxide-based disinfectant liquid.
Box including N. 6 x 1 Liter bottle and N. 6 Spray triggers.
For information on safety and waste disposal, see the safety data sheet and the technical data sheet supplied
with the unit.
The disinfectant can be used with the VIROCLEAN instrument described in the previous paragraph, or else to
directly clean and decontaminate the unit, referring to chapter 4.
Composition: 5% hydrogen peroxide, 10% ethyl alcohol.
Expiration: 2 years.
Shelf life: 12 months from when opened.