G - 4 Operator’s Manual
G EMC Guidance and Manufacturer’s Declaration
Table G-3
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
Consona N6 series is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT-
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact;
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact;
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
Transient / burst
IEC 61000-4-4
±2 kV for power
supply lines;
±1 kV for input/output
lines
±2 kV for power
supply lines;
±1 kV for input/output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±0,5 kV, ±1 kV line(s)
to line(s);
±0,5 kV, ±1 kV, ±2 kV
line(s) to earth
±0,5 kV, ±1 kV line(s)
to line(s);
±0,5 kV, ±1 kV, ±2 kV
line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, Short
interruptions and
voltage variation on
power supply input
voltage
IEC 61000-4-11
0% U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% U
T
; 1 cycle
70% U
T
for 25/30
cycle at 0°
0% U
T
; 250/300 cycle
0% U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% U
T
; 1 cycle
70% U
T
for 25/30
cycle at 0°
0% U
T
; 250/300 cycle
Mains power quality
should be that of a typical
commercial or hospital
environment. If you
require continued
operation during power
mains interruptions, it is
recommended that our
product be powered from
an uninterruptible power
supply or a battery.
Power frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE:
U
T
is the A.C. mains voltage prior to application of the test level.