G EMC Guidance and Manufacturer’s Declaration
Operator’s Manual G - 5
Table G-4
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
Consona N6 series is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Conduced RF
IEC 61000-4-6
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM
a
and amateur radio
bands between
0,15 MHz and 80
MHz
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM
and amateur radio
bands between
0,15 MHz and 80
MHz
Portable and mobile RF
communications equipment should be
used no closer to any part of system,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter. Recommended
separation distance
d = 1.2 ×
d = 2 ×
Radiated RF
IEC 61000-4-3
10 V/m
80MHz - 2.7GHz
10 V/m
80MHz - 2.7GHz
d = 1.2 × 80 MHz to 800 MHz
d = 2.3 × 800 MHz to 2.7GHz
Where, P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
b
, should be
less than the compliance level in each
frequency range
c
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.