Installation Menu System Configuration
2 - 2 0070- 0-0704-02 Passport V Operating Instructions
To access the Installation Menu:
1. Turn off the Passport V.
2. Press and hold the Discharge key on the front panel.
Simultaneously, power up the Passport V monitor.
3. Release the Discharge key when the Installation Menu is displayed.
4. Set each item as necessary using the Navigator Knob.
5. To save all of the chosen settings, choose “Save Current” before leaving this menu.
6. To return to normal operating mode, cycle power to the Passport V monitor.
The following table describes the Installation Menu structure:
MENU TITLE
ON SCREEN MENU CHOICES DEFAULT COMMENTS
Save Current Select to save current
settings as defaults.
Select Language ENGLISH
FRENCH
GERMAN
ITALIAN
SPANISH
PORTUGUESE
RUSSIAN
DUTCH
Set up at factory Select to change language.
Date Format Y/M/D
M/D/Y
D/M/Y
M/D/Y Select to change date
format.
Time Format 12 hour
24 hour
12 hour Select to change time
format.
NIBP Timeout 15 min
30 min
45 min
60 min
15 min Select to change NIBP time
out.
Temperature Units °F
°C
°F Select to change
temperature units.
Changing temperature
units will cause temperature
alarm limits to be restored
to default values.
Weight Units lbs
kg
lbs Select to change weight
units.
Height Units inches
cm
inches Select to change height
units.
CO
2
Units mmHg
%
kPa
mmHg Select to change CO
2
units. Changing CO
2
units
will cause CO
2
alarm limits
to be restored to default
values.
ECG Standard AHA
IEC
AHA Select to change standard
for ECG.
0