6 - 54 Operator’s Manual
6 Image Acquisition
3. Inject the contrast agent. Enable the timer, and save the dynamic image. Refer to “9 Contrast
Imaging”.
Data registration of same patient/body part in different phases
When different phases of CT/MR/PET data from the same patient/body part exist and one CT/MR/
PET data has been registered already, system will register the data by previous information if other
phases data are imported.
6.15.5 Acquiring/Importing Volume Data
• Only support DICOM CT/MR/PET volume data, and cine format freehand volume data.
• The following CT/MR/PET volume data not accepted:
– Slices less than 2.
– Discontinuous or corrupted data.
– Slice space not equal.
– Slice distance is 0.
– Data from varied alignment (tissue, plane).
• The liver enveloped in CT/MR/PET/freehand volume data.
• The registration effect will be better if the CT/MR/PET volume data already bears the data of
Contrast Imaging with the ultrasound image.
;
Acquiring Volume Data
• Acquiring volume data from network server
NOTE:
• It is only applied to the Ultrasound System that is configured with DICOM software
package.
• The storage path should be short and clear for network server obtaining. Space or
Chinese character is not allowed.
• Refer to “15 DICOM/HL7” to configure DICOM.
Perform the following procedure:
a. Configure DICOM service and Query/Retrieve service.
b. Load volume data to iStation.
• Acquiring volume data from external storage medium.
Perform the following procedure:
a. Save volume data to USB disk, portable hard disk, optical disk, etc.
b. Press <F2 iStation> to enter the iStation page. Select from the drop-down list of the data
source, such as USB.