17 Probes and Biopsy
Operator’s Manual 17 - 21
NOTE:
For details of the TEE probes, refer to TEE Ultrasonic Transducer manual.
17.1.8 Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and stored in a
suitable place.
• To prevent the probe from being damaged, DO NOT store it in locations where it may be
exposed to:
– Direct sunlight or X-rays
– Sudden changes in temperature
–Dust
– Excessive vibration
– Heat generators
• When the probe is sent to MINDRAY Customer Service Department or sales representative for
repair, be sure to disinfect it and keep it in the carrying case to prevent infection.
17.1.9 Sensor (Fusion Imaging) Support Cleaning and
Disinfection
Before performing the fusion imaging exam, clean and disinfect (or sterilize) the fusion sensor
support as required.
CAUTION
• When performing cleaning and disinfection of the sensor support to prevent
infection, wear sterile gloves.
• After disinfection, rinse the sensor support thoroughly with sterile water to
remove all chemical residues. Chemical residues may be harmful to the
human body.
• No cleaning and disinfecting may result in the sensor support becoming a
source of infection.
D8-2U -10°C - 50°C 10% - 90%RH (no condensation) 700 hPa - 1060 hPa
ELC13-4U -20°C - 55°C 20% - 95%RH (no condensation) 700 hPa - 1060 hPa
C6-2Gs -20°C - 55°C 20% - 95%RH (no condensation) 700 hPa - 1060 hPa
7LT4s -20°C - 60°C 10% - 95%RH (no condensation) 500 hPa - 1060 hPa
DL14-3U -10°C - 60°C 20% - 95%RH (no condensation) 700 hPa - 1060 hPa
L13-3WU -20°C - 55°C 20% - 95%RH (no condensation) 700 hPa - 1060 hPa
L15-3WU -20°C - 55°C 20% - 95%RH (no condensation) 700 hPa - 1060 hPa
Probe Ambient
temperature
Relative humidity Atmospheric
pressure

Table of Contents

Other manuals for Mindray Resona R9 Exp

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona R9 Exp and is the answer not in the manual?

Mindray Resona R9 Exp Specifications

General IconGeneral
BrandMindray
ModelResona R9 Exp
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals