7 3D/4D
Operator’s Manual 7 - 9
7.3.3 Static 3D Image Viewing
To enter/exit image viewing
•Enter: The system enters image viewing when 3D image acquisition is finished.
•Exit: Press <Freeze> or <Update> to return to image acquisition preparation status and to exit
image reviewing.
Activate MPR
Tap [A], [B], [C] or [VR] to activate sectional plane image (MPR) or 3D image (VR).
MPR Viewing
In actual display, different colors of the window box and the section line are used to identify the
section A, B and C.
Main/Sub
Render
Surface Set Surface as 3D image rendering mode.
Applicable for surface imaging, such as fetus face/hand or
foot.
NOTE:
You may have to adjust the threshold to obtain a clear
body boundary.
Max Set Max as 3D image rendering mode, displays the
maximum echo intensity in the observation direction.
This is helpful for viewing bony structures.
Min Set Min as 3D image rendering mode, displays the minimum
echo intensity in the observation direction.
This is helpful for viewing vessels and hollow structures.
X ray Set X-ray as 3D image rendering mode. Displays the average
value of all gray values in the ROI.
This is used for imaging tissues with different structure
inside or tissues with tumor.
iLive Add the light rendering effect based on the general rendering
effect.
iLive cannot be selected as sub render mode. When it is set
as main render mode, the sub render mode is disabled.
STIC Based on the movements of the fetus, rebuilt and show the anatomical structure
within a physical movement by using the interconnection between the time and
the space.
Type Parameter Description

Table of Contents

Other manuals for Mindray Resona R9 Exp

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona R9 Exp and is the answer not in the manual?

Mindray Resona R9 Exp Specifications

General IconGeneral
TypeUltrasound System
ProbesConvex, Linear, Phased Array
Imaging ModesB-Mode, M-Mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Advanced Imaging3D/4D imaging, Contrast Enhanced Ultrasound (CEUS)
ConnectivityDICOM, Ethernet, USB
PortabilityMobile system with wheels

Related product manuals