17 Probes and Biopsy
Operator’s Manual 17 - 25
17.2 Biopsy Guide
WARNING
• The person performing biopsy procedures must understand diagnostic
ultrasound thoroughly and have been trained adequately, otherwise, side
effects may be caused to the patient.
• In situations listed below, the biopsy needle may fail to penetrate the target.
The incorrect biopsy may cause various side effects in the patient.
– Use a needle-guided bracket other than that provided.
– Mount the needle-guided bracket incorrectly.
– Use a biopsy needle that is unsuitable for the type of biopsy being
performed.
– Use a biopsy needle that is unsuitable for the needle guide.
• Before and after a biopsy procedure is performed, confirm that the needle-
guided bracket is normal. Manually confirm that the parts of the needle-
guided bracket do not slip off or move from their proper positions. If the
needle-guided bracket is used when parts are not securely and correctly
installed, the patient may be injured. If an abnormality is found on the
needle-guided bracket, immediately stop using it and contact MINDRAY
Customer Service Department or sales representative.
• DO NOT use a needle-guided bracket when scanning is performed. The
needle may advance in an incorrect direction and possibly injure the
patient.
Never perform a biopsy during image scanning.
• DO NOT freeze an image while performing biopsy procedure.
• During biopsy procedures, the needle may deviate from the desired course
due to the tissue characteristics or the type of needle. In particular, needles
of small diameters may deviate to a greater degree.
• Disinfect the probe and sterilize needle-guided bracket before and after
each ultrasound-guided biopsy procedure is performed. Fail to do so may
cause the probe and the needle-guided bracket become sources of
infection.
• The needle mark displayed on the ultrasound image does not indicate the
actual position of the biopsy needle. Therefore, it should only be used as a
reference. Always monitor the relative positions of the biopsy needle during
the procedures.
• Adjust the needle mark before the biopsy procedure is performed.
• When performing biopsy procedures, use only sterile ultrasound gel that is
certified to be safe. And manage the ultrasound gel properly to ensure that
it does not become a source of infection.
• When performing the operation concerning biopsy, wear sterile gloves.
• Image of the biopsy target and the actual position of the biopsy needle:

Table of Contents

Other manuals for Mindray Resona R9 Exp

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona R9 Exp and is the answer not in the manual?

Mindray Resona R9 Exp Specifications

General IconGeneral
BrandMindray
ModelResona R9 Exp
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals