Adjustment/Conguration counter for standard scales
Purpose Automatic recording of changes to adjustment and weighing parameters using two
independent counters. The values remain saved for the life of the component.
t To display both counters, press and hold the ( key for longer than 2
seconds.
y The “Conguration counter“ is then shown in the weight display for 3 seconds
(identied by a “P“). Then the “Adjustment counter“ is displayed for another
3 seconds (identied by a “C“). After 6 seconds, the information display turns
o automatically.
Adjustment Counter Features: Counter limited to 9999
Counter at “C 0000“ for hardware commissioning
Counter cannot be reset
Counter is updated automatically when:
linearization, calibration/adjustment/preload change is successful
user calibration, adjustment or linearization weight is changed
When the following parameters are changed:
– Function J key
Zeroing range
Initial zero point range
The above parameters are reset to factory settings
Conguration Counter Features: Counter limited to 9999
Counter at “P 0000“ for hardware commissioning
Counter cannot be reset
Counter is updated automatically when:
When the following parameters are changed:
Installation location
Application lter
Stability range
– Taring
Auto zero
Weight unit 1
Weight unit 2
Weight unit 3
The above parameters are reset to factory settings
Turning the application automatic taring on/o
Restore application parameters to factory settings