DGA 9 7045899/02 EN 11.2020 Page 24
Screw the MSENSE® DGA 9 2 or 3 times clockwise by hand into the threaded flange.
Now use an M55 open-ended spanner or a 1 ½" adjustable spanner to screw the MSENSE®
DGA 9 by a further 5 or 6 turns clockwise into the threaded flange and stop when the vent screw
is in the 9 o'clock position.
Remark: If the unit can be screwed in without a great deal of effort being necessary, too little thread
sealing tape has been used. The device has to be screwed off again, the thread sealing tape removed
completely and the mounting procedure then repeated.