USER MANUAL NOVAplus
MRU GmbH, D-74172 Neckarsulm 41 / 95
6.5. Operating temperature
If the analyzer has been stored at low temperatures, it will require some
time to equilibrate to the ambient temperature before being switched
on. If it does not equilibrate, condensation will occur inside the analyzer!
If the temperature is out of its operation range you will see the follow-
ing messages on the display.
See also Chapter 11.1 Technical data, Page 79.
6.6. Controlling Condensate separator (water trap)
CAUTION
Acid from the condensate
ACID burns may result from weakly acidic liquids from
the condensate.
If you come into contact with acid, wash the area im-
mediately using a lot of water.
Control the condensate separator before and after each measure-
ment
Check that the condensate separator is empty.
Check the star filter.
Star filter white = ready for use
Star filter dark = renew
When switching off, a warning message is displayed:
Base units with optional gas cooler require a frequent check of the
pleated filter as well as the emptying of the condensate separator
ATTENTION
Once one of these messages appears you will not be
able to use the analyzer, the zeroing doesn’t start until
it has reached the specified operation temperature.