hyperion - TECHNICAL MANUAL
Cod. Meaning Description/solution
DEV -11
Calibration le error:
serial number mismatch
During sensor calibration restore: the restore le is not rela-
ted to the sensor mounted on the device
DEV -12
Calibration le error:
token mismatch
Software version not compliant with mainboard rmware
version, if the problem is persistent change the mainboard
DEV -13 Device non ready
A shot commad is received but machine is not ready. Send
a reset command and wait for ready status
DEV -14 Device non ready for message
A command is received but device is not ready. Send a re-
set command and retry.
DEV -15 No interlock
A shot commad is received but interlock is not closed. Clo-
se interlock and retry.
DEV -16 Device not in run mode
A shot commad is received but machine is not ready. Send
a reset command and wait for ready status
DEV -17 Device not in CBCT
A CBCT shot commad is received but device is not in CBCT
mode, select CBCT mode and retry.
DEV -18 Shot button pressed Release the button e retry
DEV -19 Converter not present or not ready
Verify emergency button | restart and retry| contact assi-
stance
DEV -20 Kv or ma out of range SW/FW error report to UT | restart and retry
DEV -21
Timeout changing kV, converter didn't
change the kV value on time
Restart the device and retry| contact assistance maybe is a
CAN comunication problem
DEV -22
Timeout changing mA, converter didn't
change the mA value on time
Restart the device and retry| contact assistance maybe is a
CAN comunication problem
DEV -23 Emergency button pressed Reset emergency button
DEV -24
Service command rejected, current state
does not accepts this command
Restart the device and retry
DEV -25 Loading projection error Verify Mainboard FW | change the mainboard
DEV -26
Abort Aquire Image command:
A projection is already run
SW/FW error report to UT | restart and retry
DEV -27
Y offset command received, but the device
in not in reset
Sent reset command and retry
DEV -28
Movement command rejected: device is
already moving
Restart the device and retry
DEV -29
Y +/- offset command received, but the de-
vice in not in reset
Restart the device and retry
DEV -30 Command +/- is disabled Restart the device and retry
DEV -31
Column movement rejected:
column is already moving
Software version not compliant with mainboard rmware
version
DEV -32
Column movement rejected:
column is in error ( ex: potentiometer error)
Restart the device and retry
DEV -33 Column position out of range
Software version not compliant with mainboard rmware
version
DEV -34
Set position turntable/slide/chinrest rejected:
Device is in idle
Restart the device and retry
DEV -35
Collimator moving command rejected:
Converter is not ready
Restart the device and retry
DEV -36
Collimator moving command rejected:
Converter is not ready
Restart the device and retry