159
EN
hyperion - TECHNICAL MANUAL
Cod. Meaning Description/solution
DEV -37 Converter not found
Verify emergency button | restart and retry| contact assi-
stance maybe CAN comunication problems
DEV -38
Collimator moving command rejected:
collimator is already moving
Restart the device and retry
DEV -39
Collimator moving command rejected:
CAN queue is full
Restart the device and retry
DEV -40
Command image info received but no ima-
ge is present
Software version not compliant with mainboard rmware
version
DEV -41 Image information status command rejected Restart application and device then retry
DEV -42 2D calibration command rejected Restart application and device then retry
DEV -43
Command rejected because reset routine
is in progress
Wait and retry
DEV -44
Command rejected because image down-
load routine is in progress
Wait and retry
DEV -45
Command rejected because a scan is in
progress
Wait and retry
DEV -46 CBCT eld of view not available now Restart NNT | Repeat NNT daily checks
DEV -47 CBCT exam resolution not available now Restart NNT | Repeat NNT daily checks
DEV -48 CBCT chinrest position not available now Restart NNT | Repeat NNT daily checks
DEV -49 CBCT region of interest not available now Restart NNT | Repeat NNT daily checks
DEV -50 CEPH not congured Check device conguration
DEV -51 Detector carousel not congured Check device conguration
DEV -52 Chinrest not congured Check device conguration
DEV -53 Patient size not recognized Restart application and device then retry
DEV -54 Sensor not congured Check device conguration
DEV -55 Error on reading/writing calibration data Restart application and device then retry
DEV -56 Sensor not initialized Run sensor registration and retry | Replace the sensor
DEV -57 Memory error
Restart application and device then retry | Replace the
mainboard
DEV -58 Generic PC command error Restart application and device then retry
DEV -59 Bad command from PC Restart application and device then retry
DEV -60
Parameters out of range | Xray disabled be-
cause of a previous error
Press reset to delete the error | Restart application and retry
DEV -61 Bad calibration le Check calibration le
DEV -62 Detector carousel command error Restart application and device then retry
DEV -63 Sensor slide command error Restart application and device then retry
DEV -64 Chinrest command error Restart application and device then retry
DEV -65 Device model not coherent Check device conguration
DEV -66 CBCT Fov conguration error Restart NNT | Repeat daily check | Reinstall NNT
DEV -67 Bad Brand Consol not compatible with device brand
DEV -68
Shot command rejected because cooling
down is in progress
Wait 2 minutes and retry
DEV -69 Remote RX shot command rejected Restart application and device then retry
DEV -70 Move command error
Restart application and device then retry | Check kinematic
group