hyperion - TECHNICAL MANUAL
In some cases, even if image coverage is perfect, the following message may appear:
In this case it is necessary to move the collimator manually in the indicated direction. To do this it is necessary to loosen
thefollowing2screws(Warning:thescrewatthetopiscoveredbythealuminiumlterrailwhichneedstobemoved
by acting manually on the X axis motor coupling, opening the collimator):