86
EN
hyperion - TECHNICAL MANUAL
Acquire an image and evaluate the soft tissue ramp and the auricular guides aligment:
In order to correct the alignment of the auricular guides it is necessary to adjust the position of the craniostat plate by
rotating it. This can be done after the 4 set screws that lock it has been loosened.