3.191 EDH0162En1040 06/99
MM4005 YL If variable is less
Usage IMM PGM MIP
Syntax xxYLnn
Parameters
Description xx
[int] Variable number.
nn [int] Comparison value.
Range xx 1 to 100 (integers) and 101 to 120 (floats).
nn -32767 to 32767.
Units xx
None.
nn None.
Defaults xx Missing: Error O.
Out of range: Error O.
Floating point: Error A.
nn Missing: Error C.
Out of range: Error C.
Description This command will allow execution of a command line based on a vari-
ables value. If the selected variable xx is less than the nn value, all follow-
ing commands on that line are executed. The command must be at the
beginning of a line and it applies only to that command line.
NOTE
Even though this command can be executed in immediate mode, its real
value is as a flow control instruction inside programs.
Returns
None.
Errors A Unknown message code.
C Parameter out of limits.
L Command not at the beginning of a line.
O Variable number out of range.
Rel. Commands YA Add to variable.
YE If variable is equal.
YG If variable is greater.
YS Initialize variable.
Example 5YS0 | Initialize variable #5 to 0.
5WY18 | While variable #5 is different than 18 repeat next commands.
5YL10, 2PR2.6, WS | If variable #5 is less than 10, move axis #2 incremental 2.6 units and
wait for stop.
3PR1.2, WS |
Move axis #3 incremental 1.2 units and wait for stop.
5YA1 |
Add 1 to variable #5.
WE | End while loop.