EasyManua.ls Logo

NewSonic SonoDur2 - Error Messages

Default Icon
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NewSonic SonoDur2 Issue 04 09/2016 Page 50
8.2 Error Messages
Table 8-1
Error Message
Corrective Action
Main battery is very weak, power
indicator LED is red
Charge SonoDur2
Connection is interrupted, shut-down of
Program.
Probe lost connection to instrument during operation.
Check cable and connectors. See Section 6.3 Removing
Probe Connector during Operation
COM 6: is not the virtual USB-interface,
or the Probe was not found under this
Interface. Start Program without Probe?
Probe is not connected during power on. Check correct
connections to probe. See section 6 Connect and
Disconnect Probe. SonoDur is in simulation mode.
Unknown USB device
SonoDur2 is not recognized as USB device when connected
to a PC. Check cable and firewall settings (see section
10.6.2)
Conversion into MPa solely for test loads
of equal to or higher than 100N (10kgf).
Allow anyhow?
Scale could only be used with a 100N (10kgf) probe. Select
correct probe (force), or allow conversion in Tensile
Strenghth for test loads < 100N (10kgf). In this case the
conversion is not covered by standards and do have a
higher uncertainty. Therefor the user is to decide whether
to use these values or not.
Dwell Time must be between 0 and 99
Seconds
Chose dwell time between 0 and 99 sec.
Error in handling the probe, contact time
too long.
Lift probe and start a new measurement. Check intender /
end of oscillation rod / bore of protection sleeve for
damages / contaminations. If necessary, clean surfaces
with a dry and soft cloth.
Adjustment does not fit to current Test
Force
Cal=xxxx
Accept Changes?
Adjustment was done with another test load (probe). It is
the responsibility of the inspector to use these settings or
not. (see chapter 7.7 Adjustment). If you are not sure, use
a probe with the corresponding test load.
Out of xx Scale
The measurement value is outside of the defined
conversion limits. Please choose another scale /standard or
Material-Table)
Error in probe, shift of zero frequency
out of tolerance. Reference value: xxx Hz
Actual value: yyy Hz. Program is
terminated.
Too large shift of oscillation frequency (mostly because of
contamination, drop or hard shock), program will
terminate. If the problem could not be solved due to
cleaning, please send probe back for check and
recalibration.
Your Complete Source for
Testing Equipment. Since 1969!
www.BergEng.com
Berg Engineering & Sales Company, Inc.
Tel 847-577-3980
info@bergeng.com

Table of Contents

Related product manuals