118
Chapter 4 Reprocessing Operations
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
• The alcohol used with the equipment must be 70% ethyl
alcohol or isopropyl alcohol. Using any other kind of alcohol
may result in malfunction of the equipment or the endoscope,
difficulty drying the endoscope, fire hazard, or a hazard due
to toxic vapor emitted from the alcohol.
• Alcohol is flammable and should be handled with extra care.
• Remove the alcohol in the alcohol tank and replace it with
new alcohol at least once a week. Otherwise, the alcohol in
the alcohol tank may degrade.
• Before handling the alcohol, read the precautions carefully
and use it as instructed.
Figure 4.40
E93 is displayed.
1. Confirm whether the alcohol still
present in the alcohol tank.
If no alcohol
remains in the tank.
2. Replenish the alcohol
(Refer to Section 3.6, “Inspecting and
replenishing alcohol” on page 48.)
Refer to “If the error code
[E93] is displayed again
after replenishing alcohol
and restarting the alcohol
flush process:” on page 120.
When the error code [E93] is displayed during the alcohol flush process
3. Press START button to restart the
reprocessing cycle.
4. Restarted process is completed.
Inspect the tube and connector
connected to the alcohol tank.
If any irregularity
is found.
If no irregularity is
found.
Correct
irregularities.
Press STOP/RESET
button to cancel the
reprocessing process
and perform remedial
actions to resolve the
problem.
If [E93] is displayed again.
If alcohol still present in the tank.