Chapter 4 Reprocessing Operations
123
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
9. Press the STOP/RESET button on the main control panel to stop alcohol
flushing, and then disconnect the syringe’s tube from the ventilation tube.
Figure 4.47
10. Disconnect the tube from the cap on the alcohol tank, remove the tank from
the detergent/alcohol drawer and insert the ventilation tube into the guide.
Figure 4.48