Chapter 6 Other Functions
175
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
4. Disconnect the ventilation tube from the guide of the alcohol tank.
Figure 6.25
5. Fill the syringe with air and insert the tube connected to the syringe all the
way into the far end of the alcohol tank ventilation tube.
Figure 6.26
Ventilation tube
Guide