Chapter 7 Routine Maintenance
213
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
4. Press the START button on the main panel. In about 10 seconds, the water
supply starts and the TIME/CODE display shows the remaining
reprocessing time.
Figure 7.39
5. When reprocessing is finished, the buzzer beeps and the TIME/CODE
display shows [- -] indicating that the process has ended.
Figure 7.40
6. Step on the foot pedal to open the lid.
7. Remove the gas filter case from the dome part of the lid.
Be sure to wear sterile gloves when touching the
cleaned/disinfected sampling tube to prevent contamination.
Do not touch the lid, the reprocessing basin, the container or
any area while wearing the sterile gloves. Otherwise, the
collected water may become contaminated.