Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Make sure that the distal end of the sampling tube does not
come into contact with the lid, the reprocessing basin, the
container, or any other area. Otherwise, collected water may
become contaminated.
8. Close the lid while taking the sampling tube out of the gas filter case mount
(see Figure 7.41).
Figure 7.41
9. Place a container with 2 L or large capacity under the end of the sampling
tube.
10. Press the FUNC SEL button on the subcontrol panel to select “RINSE” (see
Figure 7.42).
Figure 7.42
During rinse, hold the end of the sampling tube. Otherwise,
the water may spray out of the sampling tube, and leak from
the container.
Gas filter case mount