Chapter 7 Routine Maintenance
247
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
7. When the beeps increase in frequency, the level of disinfectant solution is
nearly full. Stop pouring the disinfectant solution and check the disinfectant
solution indicator located at the front of the equipment and slowly add more
solution (see Figure 7.82).
Figure 7.82
The MAX line and MIN line of the disinfectant solution
indicator are the reference line that indicates too much
disinfectant solution or insufficient amount of disinfectant
solution in the tank. However, the fluid level of the
disinfectant solution shown in the indicator may vary
depending on the tilt of the equipment. Therefore, error code
might not be displayed even if the fluid level in the indicator
exceeds these lines or error code might be displayed even if
the fluid level in the indicator doesn’t exceed these lines. In
this case, the equipment is not malfunctioning and can be
used. If you have any problems operating the equipment,
refer to section 4.9 “Correction of equipment tilt” in the
installation manual and correct the tilt of the equipment so
that the equipment becomes more horizontal. Then, the
indicator will indicate the fluid level more precisely.
MAX
MIN