Chapter 7 Routine Maintenance
253
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
6. To remove detergent from inside the line, connect the tube attached to the
syringe to the detergent nozzle inside the reprocessing basin. Suction the
detergent with the syringe until no more detergent comes out of the line (see
Figure 7.89).
Figure 7.89
7. Prepare a beaker or similar container and pour more than 50 ml of the
sterile water into it.
8. Put the cap in the container filled with the sterile water (see Figure 7.90).
Figure 7.90
9. To rinse the line, connect the tube attached to the syringe to the detergent
nozzle and suction more than 50 ml of sterile water with the syringe (see
Figure 7.89).
10. To remove rinse water from inside the line, connect the tube attached to the
syringe to the detergent nozzle inside the reprocessing basin. Suction the
water with the syringe until no more water comes out of the line (see Figure
7.89).
11. Prepare a beaker or similar container and pour about 30 ml of the
disinfectant solution into it.