Chapter 8 Troubleshooting and Repair
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
E81 The process
cannot be
properly
controlled.
Internal problem
with the device.
Contact Olympus.
E82 Internal
irregularity in
the device.
Irregularity in the
electrical circuitry
inside the device.
Contact Olympus.
E83 The process
cannot be
properly
controlled.
Irregularity in the
electrical circuitry
inside the device.
Contact Olympus.
E84 Malfunction of
the RFID
reader.
Irregularity in the
electrical circuitry
inside the device.
Contact Olympus.
E91 ID read error. • ID is not read.
• Multiple IDs are
read.
Press the STOP/RESET button, check that the ID detection
indicators are not lit on the main control panel, and read the IDs
again.
E92 Exceeded leak
test time.
Exceeded test
time (10 minutes).
Perform leak test again.
E93 No alcohol
remains.
• The alcohol has
run out.
• Alcohol cannot
be supplied to
the device.
• Irregularity in
the alcohol
sensor.
• Clog in the
pump or piping.
• If error occurred during the reprocessing process, refer to
“When the error code [E93] is displayed during the alcohol
flush process” on page 117.
• If error occurred during the sub-independent alcohol flush
process, check the amount of alcohol in the tank. If alcohol is
still present in the alcohol tank, refer to “If the error code
[E93] is displayed again after replenishing alcohol and
restarting the alcohol flush process:” on page 120.
E94 The equipment
is not printing.
• Printer paper
roll has run out.
• The release
lever is raised.
• Install the printer paper roll properly (follow the procedure in
Section 4.8, “Installation of the printer paper roll” in the
“Instructions-Installation Manual”).
• Check that the release lever is set in the proper position.
Error
Code
Problem Possible Causes Remedial Actions