Chapter 8 Troubleshooting and Repair
283
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Other errors and remedial action
E95 No detergent
remains.
• The detergent
has run out.
• The detergent
cannot be
supplied to the
equipment.
• Irregularity in
the detergent
sensor.
• Clog in the
pump or piping.
Refer to “When the error code [E95] is displayed during the
reprocessing process” on page 125.
Problem Possible Causes Remedial Actions
The odor of the disinfectant
solution is stronger than
expected.
• The gas filter has
expired.
• Disinfectant solution is
leaking.
• Replace the gas filter as described in Section 7.1,
“Replacing the gas filter (MAJ-822)” on page 188. If
the problem persists after replacement, contact
Olympus.
• Check if disinfectant solution is leaking. If it is, do not
use the device and contact Olympus.
Water leaks from the
equipment.
Improper installation of
water filter housing.
Stop the current process and reattach the water filter
housing as described in Section 7.2, “Replacing the
water filter (MAJ-824)” on page 192.
However, the STOP/RESET button may not work if
stopping the water might cause a hazardous situation,
for example during disinfectant solution replacement. In
this case, close the water faucet, tighten the water filter
housing, then open the water faucet again and continue
the process. If an error code is displayed, take the
appropriate remedial action for that error.
Internal problem with the
device.
Close the water faucet, set the power switch to OFF,
disconnect the power cord plug from the wall mains
outlet and contact Olympus.
Fluid leaks from the
disinfectant removal port.
Something is clogging the
disinfectant removal port.
Connect the drain connector as described in to Section
3.10, “Checking the disinfectant solution concentration
level” on page 59, push the valve on the connector
several times so that the material stuck in the port is
removed. If leakage is still detected, attach the rubber
cap to the disinfectant solution drain port and contact
Olympus.
Error