Chapter 8 Troubleshooting and Repair
285
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Disinfectant solution is
judged to be ineffective with
the test strip.
Expiration of the service life
of the disinfectant solution.
Replace the solution as described in Section 7.12,
“Replacing the disinfectant solution” on page 225.
Bacteria were detected as
a result of culture test of a
reprocessed endoscope.
Expiration of service life
of filters, degradation of
disinfectant solution, etc.
Water supply piping is not
disinfected.
Inspect the equipment as described in Chapter 3,
“Inspection Before Use”, preclean the endoscope and
put it through the reprocessing process again from the
beginning. If bacteria are detected again in the next
culture test, contact Olympus.
Bacteria were detected as
a result of culture test of
rinsing water collected from
the device.
Expiration of service life
of filters, degradation of
disinfectant solution, etc.
Water supply piping is not
disinfected.
Inspect the equipment as described in Chapter 3,
“Inspection Before Use”. If bacteria are detected again in
the next culture test, contact Olympus.
Endoscopes were not
precleaned before being
reprocessed.
Inspect the equipment as described in Chapter 3,
“Inspection Before Use”. Then, preclean the
endoscopes and reprocess them again from the
beginning.
Printed paper is not output
from the printer.
Printer paper roll has run
out.
Paper jam.
Take the remedial action by referring to Section 3.11,
“Inspecting the printer paper roll” on page 68.
Abnormal noise from the
device.
Internal problem with the
device.
Contact Olympus.
The indicator lamp of the
communication status
indicator on the main
control panel goes
out/blinks.
The connection with
external device is
incomplete or a device
malfunctions.
Check the power supply to the external devices and the
connection between the external devices and the
OER-Pro. If the problem cannot be resolved, contact
Olympus.