ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications
equipment, such as a cellular phones
Where “P” is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer, and “d” is the recommended separation
distance in meters (m).
This equipment complies with the requirements of
IEC 61326: 2002. However, if used in an electromagnetic
environment that exceeds the required maximum noise level,
electromagnetic interference may occur on this equipment.
Immunity test
IEC 61326
test level
Compliance
level
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
(150 kHz – 80 MHz)
3V (V
1
) Formula for recommended separation distance
(V
1
=3 according to the compliance level)
Radiated RF
IEC 61000-4-3
10 V/m
(80MHz 1GHz)
10 V/m (E
1
) Formula for recommended separation distance
(E
1
=10 according to the compliance level)
80 MHz – 800 MHz
800MHz 1GHz
d
3.5
V
1
-------- -
P
=
d
3.5
E
1
-------- -
P
=
d
7
E
1
-----
P
=