EasyManua.ls Logo

Olympus OER-Pro - Page 303

Olympus OER-Pro
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
297
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Recommended separation distance between portable and
mobile RF communications equipment and this equipment
Rated maximum output
power of transmitter
P (W)
Separation distance according to frequency of transmitter (m)
(Calculated as V
1=3 and E1=10)
150 kHz – 80 MHz 80 MHz – 800 MHz 800 MHz – 1 GHz
0.01 0.12 0.04 0.07
0.1 0.38 0.12 0.23
1 1.2 0.35 0.70
10 3.8 1.2 2.3
100 12 3.5 7
The guidance may not apply in some situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people. Portable
and mobile RF communications equipment such as cellular
phones should be used no closer to any part of this
equipment, including cables, than the recommended
separation distance calculated from the equation applicable
to the frequency of the transmitter.
FCC and IC information
This equipment complies with part15 of the FCC rules and the IC RSS210.
FCC ID: S8Q-GN4215
IC: 4763B-GN4215
FCC WARNING
Change or modifications not expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
d
1.2
P
=
d
0.35
P
=
d
0.7
P
=

Table of Contents