User Calibration
6.2 User Calibration Operation
6-9
Input Type and Parameter/Reference Signal
K3HB-X
M After the count stabilizes, press the U
[UP] Key.
The current value is displayed.
N Press the U [UP] Key again.
The count value is set.
Input Input type
Calibration upper limit Calibration lower limit
Parameters
Reference
signal
Parameters
Reference
signal
XVD
199.99 to 199.99 V 199. 99 199.99 V -199. 99 199.99 V
19.999 to 19.999 V 19. 999 19.999 V -19. 999 19.999 V
1.9999 to 1.9999 V 1. 9999 1.9999 V -1. 9999 1.9999 V
1.0000 to 5.0000 V 5.
0000 5.0000 V 1. 0000 1.0000 V
XVA
0.0 to 400.0 V 400.
0 400.0 V 0. 0 0.0 V
0.00 to 199.99 V 199. 99 199.99 V 0. 00 0.00 V
0.000 to 19.999 V 19.
999 19.999 V 0. 000 0.000 V
0.0000 to 1.9999 V 1.
9999 1.9999 V 0. 0000 0.0000 V
XAD
199.99 to 199.99 mA 199. 99 199.99 mA -199. 99 199.99 mA
19.999 to 19.999 mA 19. 999 19.999 mA -19. 999 19.999 mA
1.9999 to 1.9999 mA 1. 9999 1.9999 mA -1. 9999 1.9999 mA
4.000 to 20.000 mA 2.
0000 20.000 mA 4. 000 4.000 mA
XAA
0.000 to 10.000 A 10.
000 10.000 A 0. 000 0.000 A
0.0000 to 1.9999 A 1.
9999 1.9999 A 0. 0000 0.0000 A
0.00 to 199.99 mA 199.
99 199.99 mA 0. 00 0.00 mA
0.000 to 19.999 mA 19.
999 19.999 mA 0. 000 0.000 mA